The Online Residual Chlorine Analyzer is equipped with a German-made CL4.2 sensor, which adopts a membrane-covered amperometric 2-electrode system with built-in electronics. It only measures pH-dependent residual chlorine, integrates a temperature sensor for automatic temperature compensation (avoiding sudden temperature change), requires no zero-point adjustment but needs DPD-1 method calibration, and is affected by pH and O3. In terms of technical features, it has high measurement stability, easy operation and maintenance, good environmental adaptability, and flexible signal & connection.
Application
- Core Adaptation: Water bodies without surfactants (tensides) (surfactants contaminate the membrane and affect accuracy).
- Applicable Water Types: Swimming pool water, drinking water, service water, process water.
Technical Parameters
Measuring Parameter: Free chlorine; pH-dependent
Measurement Range:
Modbus RTU-Output:
0.005~2.000ppm, resolution 0.001 mg/L;
0.05~2.00ppm, resolution 0.01 mg/L;
0.5~20.0 ppm, resolution 0.1 mg/L;
4-20mA-Output:
0.005~0.500ppm, resolution 0.001 mg/L;
0.005~2.000ppm, resolution 0.001 mg/L;
0.05~5.00ppm, resolution 0.01 mg/L;
0.05~10.00ppm, resolution 0.01 mg/L;
0.05~2.00ppm, resolution 0.01 mg/L;
0.5~10.0 ppm, resolution 0.1 mg/L;
0.5~20.0 ppm, resolution 0.1 mg/L.
Response Time T90: Approximately 30 seconds