Air analyzer OxyTrans LC OEM
gasprocessfor the chemical industry

Air analyzer - OxyTrans LC OEM - Pro-Chem-Analytik GmbH & Co. KG - gas / process / for the chemical industry
Air analyzer - OxyTrans LC OEM - Pro-Chem-Analytik GmbH & Co. KG - gas / process / for the chemical industry
Add to favorites
Compare this product

Characteristics

Measured entity
gas, air
Application domain
process, industrial, for the chemical industry, for metallurgy
Configuration
for integration, compact
Technology
electrochemical
Protection level
IP65
Other characteristics
continuous, with oxygen sensor

Description

Description
The OxyTrans LC OEM is an industrial oxygen analyzer for trace and percent-level measurement. It combines an electrochemical fuel-cell sensor with a compact IP65 enclosure to deliver reliable oxygen readings in gas processes and gas production environments.

Sensor
The unit uses an electrochemical fuel-cell sensor designed for industrial use, offering high sensitivity, stable output and long service life. The sensor shows no cross-sensitivity to hydrogen and includes temperature compensation and reverse-voltage protection.

Calibration
Trace-range calibration is performed with a calibration gas, with freely selectable concentration within the measuring range. For percent-range measurements the instrument can be calibrated using ambient air.

Key features
  • Compact, space-saving design
  • Cost-effective solution for continuous monitoring
  • Long-life electrochemical sensor
  • IP65 rated enclosure for industrial environments
  • Reverse-voltage protection and temperature compensation
  • No cross sensitivity to H2


Flow-through measuring cell
  • Modular measuring cell manufactured from synthetic material for chemical resistance and easy maintenance


Standard features and options
  • Multiple sensor options available
  • Four measuring ranges (including switchable trace ranges)
  • Synthetic material cell block


Application areas
  • Nitrogen generation plants
  • Gas production and filling facilities
  • Metallurgical industry
  • Chemical industry
  • Heat treatment and bright annealing processes


Technical overview
Parameters: Oxygen %; Oxygen ppm
Method: electrochemical (fuel-cell)
Ranges: 0–25 %; 0–10,000 ppm; 0–1,000 ppm (switchable)
Accuracy: ±2 % FSD
Resolution: 0.1 ppm
Flow rate: approx. 1 l/min
Pressure: 0.1 – 1 bar
Calibration: Air (percent range) and calibration gas (trace ranges)
Response time T90: 0–10 ppm = 45 s; 0–100 ppm = 20 s
Output: 4–20 mA (2-wire)
Supply voltage: 10–35 V
Display: none
Protection class: IP65
Connection: 6 mm
Dimensions: 60 x 90
Weight: 1 kg

Technical specifications
  • Parameters: Oxygen % / Oxygen ppm
  • Method: electrochemical
  • Measuring ranges: 0–25 %; 0–10,000 ppm; 0–1,000 ppm (switchable)
  • Accuracy: ±2 % FSD
  • Resolution: 0.1 ppm
  • Response time T90: 0–10 ppm = 45 s; 0–100 ppm = 20 s
  • Flow rate: approx. 1 l/min
  • Pressure: 0.1 – 1 bar
  • Calibration: Air (percent) and calibration gas (trace)
  • Output: 4–20 mA (2-wire)
  • Supply voltage: 10–35 V
  • Display: none
  • Protection class: IP65
  • Connection: 6 mm
  • Dimensions: 60 x 90
  • Weight: 1 kg

Catalogs

No catalogs are available for this product.

See all of Pro-Chem-Analytik GmbH & Co. KG‘s catalogs

Other Pro-Chem-Analytik GmbH & Co. KG products

Products

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.