Oxygen analyzer OxyTrans LC-KF
monitoringfor ambient airindustrial

Oxygen analyzer - OxyTrans LC-KF - Pro-Chem-Analytik GmbH & Co. KG - monitoring / for ambient air / industrial
Oxygen analyzer - OxyTrans LC-KF - Pro-Chem-Analytik GmbH & Co. KG - monitoring / for ambient air / industrial
Add to favorites
Compare this product

Characteristics

Measured entity
oxygen
Application domain
monitoring, for ambient air, industrial
Measured value
concentration, trace
Technology
electrochemical
Protection level
IP65
Other characteristics
cost-effective

Description

Description
The OxyTrans LC-KF is a loop-powered 2-wire oxygen transmitter for trace and percent-level measurement. Designed for industrial process monitoring, it provides continuous oxygen concentration output via a 4–20 mA signal suitable for integration into control systems.

Sensor
The unit uses a fuel-cell type electrochemical sensor selected for sensitivity to low oxygen concentrations and long operational life. The sensor assembly meets industrial requirements for repeatability and low maintenance.

Calibration
For trace oxygen ranges, calibrate the instrument with an appropriate calibration gas at the desired concentration within the measuring range. For percent-range measurements, calibration can be performed using ambient air.

Features
  • Compact industrial transmitter
  • Cost-effective design for continuous monitoring
  • Electrochemical sensor with extended service life
  • IP65 enclosure for harsh environments
  • Reverse-voltage protection and temperature compensation
  • No cross-sensitivity to H2

Flow-through measuring cell in synthetic material
  • Modular flow cell constructed from synthetic materials to facilitate installation and maintenance

Standard Features
  • Selectable sensor types available
  • Four interchangeable measuring ranges
  • Synthetic-material cell block

Application Areas
  • Nitrogen plants
  • Gas manufacturers
  • Metallurgical industries
  • Chemical industry
  • Heat-treating and bright annealing processes

Parameters / Overview
Parameters: Oxygen %; Oxygen ppm
Method: electrochemical (fuel cell)
Range: 0 – 25 %; 0 – 10,000 ppm; 0 – 1,000 ppm (switchable)
Accuracy: ±2 % FSD
Response time T90: 0–10 ppm = 45 s; 0–100 ppm = 20 s
Output: 4–20 mA (2-wire)
Supply voltage: 10–35 V

Technical specifications
  • Parameters: Oxygen %; Oxygen ppm
  • Method: electrochemical
  • Range: 0 – 25 %; 0 – 10,000 ppm; 0 – 1,000 ppm (switchable)
  • Accuracy: ±2 % FSD
  • Resolution: 0.1 ppm
  • Flow rate: approx. 1 L/min
  • Pressure: 0.1 – 1 bar
  • Calibration: Air; calibration gas
  • Response time T90: 0–10 ppm = 45 s; 0–100 ppm = 20 s
  • Gas temperature: 0 – 50 °C
  • Output: 4–20 mA (2-wire)
  • Supply voltage: 10 – 35 V
  • Display: none
  • Protection class: IP65
  • Connection: 6 mm
  • Dimensions: 60 x 90 mm
  • Weight: 1 kg