Product IntroductionThe insulation resistance tester is used to test the electrical insulation resistance characteristics of large high-voltage transformers, electrical appliances, remote power cables, or buried cables in high interference locations such as high-voltage substations and power plants. With the rapid development of China's electric power industry, preventive experiments on electrical appliances are an important link in ensuring the safe operation and maintenance of power systems. Insulation diagnosis is an important means for detecting insulation or failure of electrical equipment. At present, electrical equipment (such as transformers, generators, etc.) is trending toward large capacity, high voltage, diverse structures, and sealing. This requires that the insulation resistance tester has the characteristics of large capacity, strong anti-interference ability, and easy to carry. To this end, our company has specially designed and developed this series of instruments for the test of electrical insulation resistance characteristics of large high-voltage transformers, electrical appliances, remote power cables or buried cables under high interference sites such as high-voltage substations and power plants.
Product ParametersRated testing voltage: 0.5kV, 1kV, 2kV, 2.5kV, 5kV, 10kV, 15kV, 20kV (8 voltage output gears in total)
Upper limit of measuring range: 100GΩ ~ 5TΩ
Lower limit of measuring range: 0.1MΩ
Output voltage error: ±5%
Short-circuit current: Approx. 5mA
Grade of accuracy: Grade 5.0
Basic error of second half measuring range: ±(5% · Rx + 1d)
Basic error of first half measuring range: ±(10% · Rx + 2d)
High voltage display error: ±(5% · Ux + 1d)
Temperature measuring error: ±0.5℃
Air humidity measuring error: ±2%RH
Power supply: 14.8V lithium battery
Service life of battery after one-time charging: 30 days, 10 times of DAR and 5 times of PI test for each day
Insulation resistance: >500MΩ (between the test line and machine case)
Voltage resistance: AC10.0kV 50Hz 1min (between the test line and machine case)
Working temperature and humidity: 0℃ ~ +40℃; Storage temperature and humidity: -20℃ ~ +50℃; Boundary dimension: 270(L) x 240(W) x 165(H) (mm)
Weight: ≈4.5kg
Product Features- Microcomputer control, menu operation, large-screen LCD dot matrix display, stable performance — intelligent instrument.
- Strong anti-interference ability; applicable in environments with severe electromagnetic interference.
- Equipped with 8 voltage output gears: 0.5kV, 1kV, 2kV, 2.5kV, 5kV, 10kV, 15kV, 20kV.
- When the high voltage is output, it can be regulated continuously.
- Automatic measurement of R15, R60, R600; automatic calculation of DAR and PI.
- Strong load capacity; short-circuit current approximately 5mA.
- Measuring range: 0 to 5TΩ, with automatic range switching.
- Combination of simulation bar pointer and digital display to show trend and accurate results.
- Testing time can be displayed at any time; buzzer reminds every 15 seconds.
- Automatic discharge of high voltage after measurement; discharge time no more than 30 seconds.
- Measures ambient temperature, air humidity and records date and time of each test automatically.
- Can save 60 groups of measuring results; data retention up to 20 years.
- RS232 serial interface for computer communication.
- Can print measuring results.
- After 5 minutes of operation the system automatically beeps and prompts shutdown.
- Comprehensive protection functions to ensure high operational reliability.
Caractéristiques / spécifications techniques- Rated testing voltage: 0.5kV, 1kV, 2kV, 2.5kV, 5kV, 10kV, 15kV, 20kV (8 gears)
- Measuring range: 0.1MΩ to 5TΩ (upper limit 100GΩ~5TΩ)
- Output voltage error: ±5%
- Short-circuit current: ≈5mA
- Accuracy: Grade 5.0
- Temperature error: ±0.5℃
- Humidity error: ±2%RH
- Power supply: 14.8V lithium battery (battery life: 30 days under specified test frequency)
- Insulation resistance (between test line and case): >500MΩ
- Voltage resistance (test line to case): AC 10.0kV 50Hz 1min
- Operating environment: 0℃~+40℃,
- Storage: -20℃~+50℃,
- Dimensions: 270 x 240 x 165 mm
- Weight: ≈4.5kg