HPLC fraction collector FRC-40

HPLC fraction collector - FRC-40 - Shimadzu France
HPLC fraction collector - FRC-40 - Shimadzu France
Add to favorites
Compare this product

Characteristics

Options
for HPLC

Description

Overview
  • Designed for preparative liquid chromatography fraction collection.
  • Space-saving design; up to six fraction collectors can be connected simultaneously.
  • Compatible with multiple rack types to match intended tube volumes.
  • Example maximum capacity: up to 3,240 fractions (6 FRC-40 units with 10 mm tubes).


FRC-10A
A proprietary band function enables accurate fraction capture even when peak elution time varies, by detecting the target component according to chromatogram variations. When connected to workstation software, fraction collection can be simulated on-screen to verify programs before run.

Drive system and collection options
  • Drive: arm-movement X-Y system for precise nozzle positioning.
  • Collection method: selectable solenoid valve or direct-through nozzle.
  • Fraction modes: manual or programmed collection by combining up to 14 parameter types (e.g., slope, level, band).
  • Number of fractions (FRC-10A): 16 to 144.
  • Compatible tubes: glass (3.5, 5, 20, 32, 50 mL) and polypropylene (4.5, 5, 25, 50 mL), or user-supplied tubes.


Additional information
FRC-10A fraction collector is required to control CBM-20A / 20Alite or SCL-10AVP.

Environmental
This product is certified as Shimadzu's Eco-Products Plus. Power consumption is reduced by approximately 59% compared with conventional Shimadzu products.

Technical specifications
  • Model designation: FRC-40 / FRC-10A
  • Application: Fraction collector for preparative liquid chromatography
  • Maximum interconnected units: up to six fraction collectors
  • Maximum fraction capacity (example): up to 3,240 fractions (6 FRC-40s, 10 mm tubes)
  • FRC-10A band function: proprietary band detection to capture target components despite peak elution variability
  • Simulation: fraction simulation available when connected to workstation software
  • Drive system (FRC-10A): arm-movement X-Y system; selectable solenoid valve or direct-through nozzle
  • Fraction modes (FRC-10A): manual or programmed using up to 14 parameter types
  • Number of fractions (FRC-10A): 16–144
  • Compatible collection tubes: glass 3.5 / 5 / 20 / 32 / 50 mL; polypropylene 4.5 / 5 / 25 / 50 mL; user-supplied allowed
  • Control requirement: FRC-10A required to control CBM-20A / 20Alite or SCL-10AVP
  • Eco certification: Shimadzu Eco-Products Plus (approx. 59% lower power consumption vs conventional products)