OVERVIEW The ADT 780 series builds on the experience from the ADT 761 calibrators. The ADT780S laboratory pressure controller offers advanced functions with selectable measurement ranges (70 bar or 200 bar). Each unit is supplied with a preselected sensor from the available range to suit the application. Internal and external sensor modules extend measurement coverage while maintaining high accuracy.
Modular design Each unit is equipped with a selectable sensor from the catalogue. Standard sensors (ADT 160A-CP-XXX) use a silicone diaphragm and provide one-year accuracy of 0.02% full scale. For higher performance, aged quartz sensors deliver 0.005% reading + 0.005% full scale. After assembly every sensor is tested and compensated over 0 to 50°C. ADT780S accepts one internal and one external sensor, interchangeable via intelligent pressure modules. Every ADT780 includes an internal barometric sensor for relative or absolute calibrations.
Autonomous solution ADT780S units can be paired with an electric pump (EPUMP) generating up to 70 bar; a gas booster enables operation up to 200 bar. This configuration eliminates the need for a permanent gas bottle and improves solution mobility.
ADT780-1K – EPUMP (electric pump) Generates vacuum to 70 bar; integrated filter and liquid trap; pressure range -0.95 to 75 bar; mass 31.6 kg; dimensions 440 x 235 x 550 mm; power 100–250 V; medium: air; pressure connection port G1/8F; storage temperature -20°C to 70°C; operating environment 0–95% RH non-condensing; bench mounting (rack kit optional); pressure limit control via 2 LEDs; time to reach max pressure: approx. 2 minutes with ADT 780 controller.
Specifications (summary) - Measurement ranges: ADT 780S-1K: -0.95 to 70 bar (-13.5 to 1,000 psi); ADT 780S-3K: -0.95 to 200 bar (-13.5 to 3,000 psi).
- Control stability: 0.003% FS (stability based on % span of control range).
- One-year stability: standard sensor 0.01% FS; upgraded quartz sensor 0.005% FS.
- Media: clean gases.
- Overpressure indication: 103% to 120% depending on sensor.
- Resolution: selectable 4 to 7 digits.
- Pressure types: relative and absolute.
- Warm-up time: 15 minutes.
- Pressure settling time: 20 s (10% span change, 50 ml test volume).
- Units selectable: Pa, hPa, kPa, MPa, bar, mbar, psi, mmHg@0°C, inHg@0°C, inH2O@4°C, mmH2O@20°C, kg/cm2, torr, atm, etc.
- Minimum control pressure: ADT 780S-1K: 0.0001 psi; ADT 780S-3K: 0.001 psi.
Pressure ranges — Absolute sensors (extract) - AP15R: 0 to 1.0 bar, media G,L, accuracy 0.01% span, overpressure 2X/1.2X, suggested ADT780-1K, internal/external.
- AP30R: 0 to 2.0 bar, media G,L, accuracy 0.01% or 0.003% PE (option), overpressure 2X/1.2X, ADT780-1K, internal/external.
- AP1KR: 0 to 71 bar, media G,L, accuracy 0.01% or 0.003% PE, overpressure 2X/1.2X, ADT780-1K/3K, internal/external.
- AP2KQ/AP3KQ: up to 141/201 bar, gas media, accuracy 0.005% reading + 0.005% FS, overpressure 3X/1.1X, ADT780-3K, internal/external.
Pressure ranges — Gauge/relative sensors (extract) - CP10: ±0.7 bar g, accuracy 0.02% FS, overpressure 3X/1.2X, ADT780-1K, internal/external.
- CP1K/CP3K: up to 70/200 bar g, accuracy 0.02% FS, overpressure 3X/1.2X, ADT780-1K/3K as suggested, internal/external.
- CP5K/CP10K: higher ranges (up to 700 bar), external sensor only.
Technical specifications - Power: 100–250 V, 50/60 Hz.
- Pressure connector: G1/8 F.
- Storage: -20°C to 70°C; Operation: 0–95% RH non-condensing.
- Display: 7" color touchscreen.
- Mass: 15 kg; Dimensions: 419 x 440 x 132 mm.
- Mounting: bench (rack kit optional).
- Shock/vibration: 4G shock; 1G 10Hz–500Hz vibration.
Other features - Connectivity: RS232, USB, LAN, WiFi (802.11 b/g/n).
- Analog outputs selectable: 0–10 V, 0–5 V, 0–20 mA, 4–20 mA.
- Operating modes: control, measure, venting.
- Interface languages: English, German, Spanish, French, Italian, Portuguese, Russian, Japanese, Chinese.
- Traceable NIST certificate included as standard; CE compliant; 1-year warranty.
Models and accessories supplied/optional - Models: ADT 780S-1K (70 bar), ADT 780S-3K (200 bar).
- Supplied accessories: external manifold ADT 127 with connection cable, mains adapter 110/220 V, connection cable for external pressure modules, pressure adapter set, NIST traceable certificate, O-RINGs (20 pcs).
- Optional accessories: external pressure modules ADT 160A, manifold ADT 121-X for 200 bar, 200 bar booster, 70 bar electric pump (EPUMP), vacuum pump, RS232-USB adapter, liquid trap 9907-780, transport case, calibration software (PCAL, ACAL), rack mounting kit.