Product description: The CIT06-10S Oxidizing Liquid Tester measures the propensity of a liquid to increase the burning rate or intensity of a combustible substance when thoroughly mixed, in accordance with UN Recommendations on the Transport of Dangerous Goods — Tests and Criteria Manual (34.4.2) and GB/T 21620-2008. The instrument places the pressure vessel inside a detection system with an explosion-proof observation window on the front door; during the test it accurately records internal pressure and time and automatically plots the pressure curve.
Product parameters:- Working environment: 5 ℃ ~ 30 ℃, <85% RH
- Pressure sensor response capability: ≤5 ms for 0–3 MPa pressure changes
- Pressure measuring range: 0–3 MPa
- Pressure measurement accuracy: ±0.25% FS
- Heating: nickel/chromium wire
- Disk burst pressure: 2.5 MPa ±5%
- Standard time range: 0–60 s
- Timing accuracy: ±1 ms
Product Features:- Embedded processor with Windows CE operating system
- 8-inch LCD touchscreen displaying test status in real time
- Automatic pressure vessel tightness detection (initial pressure drop detection)
- Ignition sound and light alarm
- Stainless-steel seamless welded tank and pressure vessel; corrosion-, high-temperature- and high-pressure-resistant pressure sensor
- Automatic timing and precision pressure monitoring system with real-time detection and display of pressure variations
- Automatic pressure relief device: disc puncture for safe pressure release after test
- Automatic drawing and storage of pressure curves with raw data view
- Reaction residue collection and splash protection to avoid sample contamination outside the vessel
- Remote ignition to ensure operator safety
- Fully automated test workflow with parameter setting, data storage and report viewing
Technical specifications:- Model: CIT06-10S
- Working environment: 5 ℃ ~ 30 ℃, <85% RH
- Pressure measuring range: 0–3 MPa
- Pressure measurement accuracy: ±0.25% FS
- Pressure sensor response: ≤5 ms for 0–3 MPa changes
- Disk burst pressure: 2.5 MPa ±5%
- Heating method: nickel/chromium wire
- Standard timing range: 0–60 s; timing accuracy ±1 ms
- Interface: 8-inch LCD touchscreen (real-time status)
- Safety features: explosion-proof front window, ignition sound and light alarm, automatic pressure relief, remote ignition, residue collection and splash protection
- Data capabilities: automatic pressure curve drawing, raw data storage and report viewing