Pressure test unit CIT06-20S

Pressure test unit - CIT06-20S - Labfreez Instruments Group Co., Ltd
Pressure test unit - CIT06-20S - Labfreez Instruments Group Co., Ltd
Pressure test unit - CIT06-20S - Labfreez Instruments Group Co., Ltd - image - 2
Pressure test unit - CIT06-20S - Labfreez Instruments Group Co., Ltd - image - 3
Add to favorites
Compare this product

Description

Product overview
CIT06-20S oxidizing liquid and time/pressure tester is designed to determine the oxidizing potential of liquid substances and to perform enclosed time/pressure ignition tests to assess deflagration risk under package-relevant pressures. The instrument records pressure–time curves and evaluates whether a liquid increases burning intensity or poses a hazard during transport and handling.

Standards compliance
The CIT06-20S complies with the UN Recommendations on the Transport of Dangerous Goods, Manual of Tests and Criteria: Test O.2 (oxidizing liquids) and the Time/Pressure test methods referenced in 11.6.1 Test 1(c)(1), 12.6.1 Test 2(c)(1) and 23.4.1 Test C.1.

Product features
  • Embedded processor running Windows CE for stable test control
  • 8-inch LCD touchscreen with real-time status and data display
  • Automatic leak detection via initial pressure drop rate monitoring
  • Electronic remote-controlled ignition with audible and visual alarms
  • Seamless-welded stainless-steel inner tank and corrosion/high-temperature resistant pressure vessel and sensor
  • High-precision, fast-response pressure detection and automatic timing
  • Automatic pressure relief via burst disk after test
  • Automatic pressure-curve plotting, data storage and report viewing
  • Residue collection and splash protection to prevent external contamination
  • Fully automated test sequence with parameter setting and remote operation


Product parameters
  • Model: CIT06-20S
  • Pressure measuring range: 0–5 MPa
  • Pressure sensor response: ≤5 ms for 0–3 MPa lumen changes
  • Pressure accuracy: ±0.1% FS
  • Burst disk pressure: 2.5 MPa ±5%
  • Ignition: Electronic ignition with remote-control capability
  • Standard time range: 0–60 s; timing accuracy ±1 ms
  • Working environment: -5 °C to 45 °C, <95% RH