Product descriptionThe CIT10-10S enclosed-space aerosol ignition tester evaluates aerosol flammability in a closed or restricted volume in accordance with UN 'Recommendations on the Transport of Dangerous Goods — Manual of Tests and Criteria' (section 31.5) and GB/T 21631-2008. The instrument reproduces confined-space conditions to determine whether ignition, flash or explosion occurs when an aerosol is discharged.
Instrument overviewThe system comprises a display/control unit and a detection module. The detection module is a cylindrical vessel with integrated optical sensors and a high-precision thermocouple to detect ignition or knocking events. The display/control unit features an LCD for real-time monitoring and test control. Aerosol cans are positioned and actuated automatically; the device supports vertical and inverted orientations.
Product parameters- Operating environment: 5 ℃ to 30 ℃, <85% RH
- Ambient temperature measurement range: -10 ℃ to +60 ℃; accuracy ±0.5 ℃
- Humidity range: 5% to 95% RH; accuracy ±3.0% RH
- Combustion detection response time: ≤6 ms
- Differential pressure measurement range: -5 to 5 kPa; accuracy ±0.1% FS
- Timing accuracy: ±0.1 s
- Ignition source: standard candle flame, adjustable pedestal height 20 mm
- Cylindrical vessel dimensions: Φ600 mm × 720 mm
- Aerosol can handling: automatic flip for orientation
- Compatible aerosol can size: height 120–400 mm, diameter 30–70 mm
- Nozzle-to-vessel distance: adjustable 30–35 mm
- Effusion plate: removable collector for discharged liquid
Product features- Embedded processor with Windows CE
- 8-inch LCD touch screen for real-time status and control
- Two operation modes: Standard and Custom
- Automatic spray actuation and clamping; supports upright and inverted cans
- Dual detection: differential pressure sensors and nanosecond light sensors to capture flame or explosion events
- Integrated ventilation to exhaust smoke from the vessel for test accuracy and operator safety
- Fully automated test sequence with parameter setting, data storage and report viewing
- Temperature and humidity module for real-time environment monitoring
Technical specifications- Model: CIT10-10S
- Cylindrical vessel: Φ600 mm × 720 mm
- Working environment: 5 ℃ to 30 ℃, <85% RH
- Ambient temperature range: -10 ℃ to +60 ℃; accuracy ±0.5 ℃
- Humidity range: 5% to 95% RH; accuracy ±3.0% RH
- Pressure (differential) range: -5 to 5 kPa; accuracy ±0.1% FS
- Combustion detection response time: ≤6 ms
- Timing accuracy: ±0.1 s
- Compatible aerosol can size: height 120–400 mm, diameter 30–70 mm
- Nozzle distance adjustable: 30–35 mm
- User interface: 8-inch LCD touch screen running Windows CE