Product introduction:DSC is designed to determine heat transitions related to temperature and heat flow. It is widely used in polymer development, performance testing and quality control. Typical analyses include glass transition, melting point, cold crystallization, crystallization, phase transition and oxidation induction time (OIT).
Main Features:- New oven design ensures high resolution and a stable baseline
- Digital air flow meter provides accurate flow control; test data can be recorded directly into the database
- Dual control: instrument operation from the mainframe or via software, with a user-friendly interface
Technical parameters (full list):- DSC range: 0 ~ ±500 mW
- Temperature range: room temperature ~ 800 °C (air-cooled)
- Temperature range (options): -50 °C ~ 800 °C (semiconductor refrigeration) *
- Temperature range (options): -100 °C ~ 800 °C (liquid nitrogen refrigeration) *
- Heating rate: 1 ~ 80 °C / min
- Cooling rate *: 1 ~ 20 °C / min
- Temperature resolution: 0.1 °C
- Temperature fluctuations: ±0.1 °C
- Temperature repeatability: ±0.1 °C
- DSC noise: 0.01 μW
- DSC resolution: 0.01 μW
- DSC accuracy: 0.1 μW
- DSC sensitivity: 0.1 μW
- Control Mode: Rising temperature, constant temperature, cooling temperature (fully automatic programmed control)
- Curve scanning: Rising scan, *cooling scan
- Atmosphere control: Embedded digital flow meter & software control
- Color / Display: 24 bit 7 inches LCD touch screen display
- Data interface: RS232 interface
- Parameter standard: equipped with standard material (indium, tin, and lead) for temperature and enthalpy correction
- Warranty: Five years
- Note: * for selectable options; technical indicators can be adjusted according to customer requirements
Characteristics / technical specifications:- Model / SKU: DSC100 L
- DSC range: 0 ~ ±500 mW
- Temperature range: RT ~ 800 °C (air-cooled); optional -50 °C or -100 °C to 800 °C with refrigeration
- Heating rate: 1 ~ 80 °C/min
- Cooling rate (optional): 1 ~ 20 °C/min
- Temperature resolution / fluctuation / repeatability: 0.1 °C; fluctuation ±0.1 °C; repeatability ±0.1 °C
- DSC noise / resolution: 0.01 μW; accuracy / sensitivity: 0.1 μW
- Control modes: programmed automatic control (heating, hold, cooling)
- Atmosphere control: digital flow meter and software control
- Interface: RS232
- Standard calibration materials: indium, tin, lead (for temperature and enthalpy correction)
- Warranty: Five years
- Display: 7 in LCD touch screen